RM RM Eye Manuál s instrukcemi Strana 81

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 92
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 80
79
SPECIFICATIONS AND PERFORMANCE
Guidance and manufacturer's declaration - electromagnetic immunity
The RM-8900,KR-8900 is intended for use in the electromagnetic environment specified below.
The customer or the user of the RM-8900,KR-8900 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150kHz to
80MHz
3 V/m
80MHz to 2,
5GHz
3 V
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the RM-8900,KR-8900, including
cables, than the recommended separation
distance calculated from the equation appli-
cable to the frequency of the transmitter.
Recommended separation distance
d = 1.2
d = 1.2 80MHz to 800MHz
d = 2.3 800MHz to 2, 5GHz
where P is the maximum output power rat-
ing of the transmitter in watts (W) accord-
ing to the transmitter manufacturer and d
is the recommended separation distance
in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the compli-
ance level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1
NOTE 2
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
b
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be con-
sidered. If the measured field strength in the location in which the RM-8900,KR-8900 is
used exceeds the applicable RF compliance level above, the RM-8900,KR-8900 should
be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the RM-8900,KR-8900.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
P
P
P
Zobrazit stránku 80
1 2 ... 76 77 78 79 80 81 82 83 84 85 86 ... 91 92

Komentáře k této Příručce

Žádné komentáře